Towards Sustainable Magnetic Resonance Imaging: Insights from long-term, high-resolution energy recordings across an entire scanner fleet

📅 2026-09-21
📈 Citations: 0
✨ Influential: 0
📄 PDF
🤖 AI Summary
研究通过长期高分辨率记录整个MRI机队的能耗,分析了影响MRI能耗的因素,发现扫描时长比平均功率更显著影响能耗。
📝 Abstract
Magnetic resonance imaging (MRI) is among the most energy-intensive diagnostic modalities in healthcare, yet its energy consumption and the factors influencing it remain insufficiently understood. This study aims to establish a comprehensive baseline of MRI energy consumption by characterizing energy demand across a scanner fleet, examining scanner utilization and operating patterns in clinical practice. Concurrently, it investigates the relationships between energy consumption and relevant operational and acquisition features. On average, a single MRI measurement consumed 0.43 kWh, while a complete examination consumed 13.50 kWh. In general, substantial differences in energy consumption were observed between MRI scanners and their corresponding operating modes (scan, idle, and eco-power mode). These variations may be related to differences in scanner operating patterns, employed examination protocols, and their resulting acquisition parameters. Idle and eco-power modes accounted for more cumulative energy consumption than active scanning. However, these energy shares should always be interpreted in relation to scanner occupancy, as utilization patterns strongly influence the distribution of energy across those operating modes. Lastly, linear regression analysis showed that energy consumption was more strongly associated with scan duration than with average power, suggesting that scan duration may be an important factor influencing MRI energy consumption. ...
Problem

Research questions and friction points this paper is trying to address.

Magnetic Resonance Imaging
Energy Consumption
Scanner Fleet
Innovation

Methods, ideas, or system contributions that make the work stand out.

energy consumption
operating modes
scan duration
high-resolution recording
scanner fleet
🔎 Similar Papers
F
Florian Leonhard Raab
Medical Image and Data Analysis (MIDAS.lab), Tübingen University Hospital
F
Fiona Mankertz
Department of Diagnostic and Interventional Radiology, Tübingen University Hospital
N
Nour Maalouf
Department of Diagnostic and Interventional Radiology, Tübingen University Hospital
J
Josephine Berger
Department of Diagnostic and Interventional Radiology, Tübingen University Hospital
A
Andreas Lingg
Department of Diagnostic and Interventional Radiology, Tübingen University Hospital
R
Reza Dehdab
Department of Diagnostic and Interventional Radiology, Tübingen University Hospital
Sebastian Werner
Sebastian Werner
Researcher, TU Berlin
Distributed SystemsCloud Native Application ArchitecturesSustainable Cloud ComputingISE@TUB
J
Judith Herrmann
Department of Diagnostic and Interventional Radiology, Tübingen University Hospital
A
Andreas Brendlin
Department of Diagnostic and Interventional Radiology, Tübingen University Hospital
S
Sebastian Gassenmaier
Department of Diagnostic and Interventional Radiology, Tübingen University Hospital
S
Suhas Siddaramu
Medical Image and Data Analysis (MIDAS.lab), Tübingen University Hospital
Fabian Wagner
Fabian Wagner
Siemens Healthineers
machine learningknown operator learningimage processingoptics
J
Julian Wohlers
Magnetic Resonance, Siemens Healthineers AG
S
Shreeja Varadarajan
Magnetic Resonance, Siemens Healthineers AG
G
Gurlal Singh
Magnetic Resonance, Siemens Healthineers AG
J
Jens Gühring
Magnetic Resonance, Siemens Healthineers AG
R
Rainer Schneider
Magnetic Resonance, Siemens Healthineers AG
Konstantin Nikolaou
Konstantin Nikolaou
University of Tuebingen, Department of Radiology
Radiology
S
Saif Afat
Department of Diagnostic and Interventional Radiology, Tübingen University Hospital
Thomas Küstner
Thomas Küstner
University Hospital of Tübingen
Medical Signal ProcessingMagnetic Resonance ImagingMachine LearningMotion CorrectionMedical Image Analysis